BioNomadix Wireless RSP with ECG Amplifier

Request More Info
BioNomadix Wireless Respiration with ECG Set
Part #: BN-RSPEC

The Dual Wireless Respiration and ECG BioNomadix module pair consists of a matched transmitter and receiver pair optimized for respiration and electrocardiogram data; record one or both signals.

  • RSP (CH A)–measure abdominal or thoracic expansion and contraction while breathing
  • ECG (CH B)–record electrical activity generated by the heart

The high signal to noise ratio and high time-based sampling resolution permit the pair to be used for precise respiratory studies and exacting heart rate variability studies. Physiological signal data is transmitted at a rate of 2,000 Hz, providing high resolution waveforms at the receiver module’s output.

The largest ECG signal which can be measured is 10 mV P-P, the smallest is less than 1 µV rms. Raw ECG data is bandlimited from 0.05 Hz to 150 Hz, to provide a very high quality recording. Biopotential amplifier/transmitter signal to noise ratio is very high (in excess of 70 dB), as the noise floor for ECG signal transmission is only 0.9 µV rms (bandwidth 0.05-150 Hz).

These units interface with the MP160 (or older MP150) data acquisition and analysis platform and AcqKnowledge software, allowing advanced analysis for multiple applications and supporting acquisition of a broad range of signals and measurements.

See More...

MODULAR CONSTRUCTION

Amplifiers snap together for easy system configuration and re-configuration.

Intuitive, Elegant AcqKnowledge Software

Powerful automated analysis. Instantly & easily view, measure, analyze, transform, and report data.

Powerful MP160 Data Acquisition and Analysis System

Flexible, proven modular data acquisition and analysis system for life science research.

BioNomadix Wireless Respiration with ECG
View Spec PDF

Are You Looking for These?

Part #: BN-RSPEC
Categories: Amplifiers - Research, BioNomadix Wireless Physiology - Research
Subcategories: Biopotential Amplifiers - Research, Transducer Amplifiers - Research, Transmitter-Receiver Amplifiers - Research

Alternatives
For the same Platform

Equivalents
With a different Platform

Details

Compatibility


The module pair is designed to emulate a “wired” connection from subject computer, with all the benefits of a fully-wireless recording system.

RSP Signal

Specifically designed to measure respiration activity associated with circumferential changes in the thoracic or abdominal regions of the torso. Attach the transmitter near or on the subject’s torso via the stretchy Velcro strap (included) or use the pocketed, stretch-mesh BioNomadix Shirt and connect the respiration transducer(s).

  • As the subject inspires, the thoracic and/or abdominal circumference expand slightly, increasing the output signal level.
  • As the subject expires, the circumference reduces, dropping the output signal level.

Raw RSP data from the module pair is bandlimited from DC to 10 Hz, to provide for the measurement of relatively static respiratory conditions, such as cessation of breathing, up to the extremely rapid respiratory effort variations (up to 600 breaths per minute) associated with coughing or sneezing. Additional internal highpass and lowpass optional filters provide for high quality amplification of respiratory effort signals even during periods of vigorous exercise or activity.

Use AcqKnowledge to evaluate the difference between thoracic and abdominal respiratory effort. The respiratory effort transducer is sensitive enough to measure the circumferential change associated with each heart beat, so AcqKnowledge can be used to extract both heart rate BPM and breathing rate BPM from respiratory effort signals. Using AcqKnowledge, both fast or slow response respiration data can be isolated to accommodate high quality signal component identification. Respiratory measurement can also be synchronized to a specific stimulus for stimulus/response investigations, such as those associated with challenge and threat.

ECG Signal

  • Record with a 3-lead set (BN-EL15-LEAD3, BN-EL30-LEAD3, or BN-EL45-LEAD3) and disposable electrodes (EL504 recommended)

The module pair has a fixed gain of 2,000 for the ECG signal. Using AcqKnowledge, wide bandwidth ECG data can be filtered to lower values to accommodate high quality ECG signal isolation even in the presence of extensive subject motion.

AcqKnowledge software automatically scores the data and extracts the measurements of interest on a cycle-by-cycle basis. The results are automatically exported to Excel or pasted in the Journal file. AcqKnowledge also includes a fully automated HRV analysis feature. The HRV analysis provides values for VLF, LF, HF, VHF, sympathetic, and vagal, as well as the sympathetic / vagal balance.

ECG100C ECG amplifier and AcqKnowledge automated ECG analysis ECG analysis 3d surface plot (waterfall) of Lead II ECG recording

 


BioNomadix are configured for specific signal types, so you don’t have toadjust settings on the hardware or software for high quality data.

Multiple BioNomadix can be used to provide the optimum signal combination for a research study requiring biopotential measurements from multiple subjects or multiple locations.

Videos

Integrating BIOPAC Research Systems | Data Acquisition & Analysis

BIOPAC Research Solutions | ECG

HRV Analysis: Fully-automated Routine in AcqKnowledge

RSA Analysis

Automated ECG Analysis- Lead II

Classify Heartbeats

AcqKnowledge Waterfall Plot

Support

Application Notes

Applications

Downloads/Resources

Knowledge Base

Recommended Items

Spotlight On
free BIOPAC webinar

HRV | Ultra-Short-Term & Short-Term

In this webinar, we will demonstrate recording of high-quality ECG data as well as preparing data for analysis through UST HRV measurements during acute stress, Valsalva maneuver and paced breathing.

You Will Learn
– Physiological basis of HRV
– Impact of measurement durations on HRV
– Recording good quality ECG & preparing data for analysis
– Analysis of short and ultra-short duration HRV
– Single-epoch and multi-epoch analysis
– Focus areas and event-based analysis
– Automation with scripting and workflow

Watch On Demand

Register Now
Latest News

New Citations | BIOPAC’s AcqKnowledge Software

BIOPAC tools are used for studies in a variety of different fields and disciplines. AcqKnowledge software is one of our universally applicable tools, used by thousands of labs around the world and cited in thousands of peer reviewed publications, such as the following studies… The psychophysiology of guilt in healthy adults. Cognitive, Affective, & Behavioral Neuroscience, 1-18. […]

Electrocardiography Guide Available

BIOPAC’s comprehensive Introductory ECG Guide addresses fundamental to advanced concerns to optimize electrocardiography data recording and analysis. Topics include: ECG Complex; Electrical and Mechanical Sequence of a Heartbeat; Systole and Diastole; Configurations for Lead I, Lead II, Lead III, 6-lead ECG, 12-lead ECG, precordial leads; Ventricular Late Potentials (VLPs); ECG Measurement Tools; Automated Analysis Routines for extracting, […]

Read All
Request a Demonstration
Request a Demonstration