BIOPAC® Systems, Inc. Logo

Grounding guidelines

When a single subject is connected via wires to a single MP system, there should generally be one and only one ground connection to that subject*. When multiple subject grounds are used, ground loops can occur: current will flow between the different grounds. This will result in distorted data.

When recording with the EDA100C (GSR100C), EBI100C, or the NICO100C, a ground is already provided through the VIn- or I IN connection (in these three cases as well as their -MRI brethren, an internal connection makes VIn- for EDA amplifiers or I IN for the bioimpedance amplifiers electrically identical to ground). Consequently, no other ground connection to the subject* need be used. In some cases, particularly when power line noise is apparent despite the presence of a single ground, it is beneficial to have more than one ground.  In the case of the EDA100C (or GSR100C), additional connections can be made through AC-coupled lead adapters (CBL205).

If using BioNomadix wireless transmitters, the subject requires one ground per transmitter. This only applies to units making an electrical connection with the subject. In addition, the BN-PPGED does not require an extra ground, as the VIN- of the EDA already acts as a ground. However, BN-ECG2, BN-EMG2, BN-EEG2, BN-EOG2 and BN-EGG2 all require separate ground connections.

* “To the subject” meaning specifically an electrical connection to the subject.  Many other BIOPAC components, such as the respiratory effort transducer and photoplethysmogram transducer have ground connections to the equipment but are not relevant to this topic because these devices do not make electrical connections to the subject and thus will not produce ground loops through the subject.

To record EDA (GSR) simultaneously with other biopotential signals (for instance, ECG, EEG, EOG, EGG, EMG, or ERS):

1. Connect the Vin- lead of the EDA as ground (when using disposable electrodes there will be nothing connected to the “GND” connection on the front of the GSR100C).

2. Connect an AC-coupled ground to the ECG100C (e.g., LEAD110A connected through CBL205 to the ECG100C’s “GND” connection).

The AC coupled lead will not influence the EDA measurement when all electrodes are connected, but it will continue to function as a ground for the ECG amplifier if the electrode connected to VIn- on the GSR100C is disconnected from the subject.

For more on this topic see “Special Cases” under “Multiple amplifiers per subject or multiple subjects per system.”

Note—If using any two of the following on the same subject connected directly to the same MP system at the same time: EDA100C (GSR100C), EBI100C, or NICO100C, ground loops will be a problem; you will not be able to make absolute measurements (such as SCL). BIOPAC recommends the use of BioNomadix modules in this case as these transmitters do not share a common ground. Alternatively, one or more of the measures may be made through an additional MP system or with an IPS100C/HLT100C/OUTISO to optically isolate the ground connections.  More information may be found under Galvanic Isolation Guidelines.

Associated Applications

Spotlight On
Ventilator Validation Kit

Ventilator Validation and Test Equipment

The Ventilator Validation Kit supports new prototype development, existing design validation for production testing, mobile field testing and verification, and other uses where accurate pressure and flow measurements are critical. The Ventilator Validation Kit provides ventilator manufacturers and prototype developers with a complete solution for validating new medical/hospital-grade ventilator products intended for human use. The VVK100-SYS includes […]

View All
Latest News

Researchers Respond | iSAVE: Rapidly deployable system to expand ventilator capacity

MIT and Harvard researchers developed a new technique to expand ventilator capacity Individualized System for Augmenting Ventilator Efficacy (iSAVE): A Rapidly deployable system to expand ventilator capacity “The iSAVE is a rapidly deployable solution in response to the urgent shortage of mechanical ventilators for respiratory support during the COVID-19 pandemic. In contrast to previously described […]

New Ventilator Validation Kit for Rapid Development & Field Testing

New Ventilator Validation System Allows Device Manufacturers to Test Ventilator Pressure and Volume System supports rapid development and testing of hardware for COVID-19 treatment Goleta, California – March 24, 2020 BIOPAC Systems, Inc. announced a new Ventilator Validation System for engineering and device manufacturers to use when designing and testing new and existing ventilators for […]

Read All
Request a Demonstration
Request a Demonstration