Electrodermal activity measurements

Definitions:

 
EDA = electrodermal activity
SRR = skin resistance response

SRL = skin resistance level

SCR = skin conductance response

SCL = skin conductance level

SPR = skin potential response

SPL = skin potential level

 

SRR and SCR are reciprocals of each other but are otherwise the same thing – discrete changes in skin resistance/conductance. To measure ONLY these changes, data may be acquired through a high pass filter.  The EDA100C (older model: GSR100C) has 0.05Hz and 0.5Hz HP filters which may be used for this purpose.

 

SRL and SCL are similarly reciprocols but otherwise the same thing – the baseline skin resistance/conductance. To record this form of activity, data must not be high pass filtered.  EDA100C (GSR100C) must be set up with both HP filter switches in the “DC” positions. 

 

NOTE: Recording DC signals does NOT prevent the recording of SCRs.  However, limiting the acquired data by passing it through a high pass filter for the purpose of recording ONLY SCRs prevents the recording of tonic SCL.  To take advantage of all EDA analysis features in AcqKnowledge or, more generally, to fully characterize EDA, SCL must be recorded, so HP filters should both be set to DC. 

 

While SRR, SCR, SRL and SCL rely on an external source of current (they are exosomatic), SPR and SPL do not (they are endosomatic). The EDA (GSR) series amplifiers are not appropriate for measuring SPR and SPL. (If you wish to work with these measures please contact support at BIOPAC for details.)

 

Skin conductance is usually measured in micromhos. Conductance is the reciprocal of resistance. 1 micromho of SCL is equivalent to 1,000,000 ohms SRL. The siemens and mho (as well as units derived from them such as microsiemens and micromho) are different names for the exact same unit.

 
Source: Psychophysiology Human Behavior and Physiological Response Fourth Edition
John L. Andreassi, Baruch College City University of New York,

LAWRENCE ERLBAUM ASSOCIATES, PUBLISHERS, 2000 Mahwah, New Jersey London

 

Page last modified 24Dec2014

Associated Applications

Associated Application Notes

Spotlight On
Ventilator Validation Kit

Ventilator Validation and Test Equipment

The Ventilator Validation Kit supports new prototype development, existing design validation for production testing, mobile field testing and verification, and other uses where accurate pressure and flow measurements are critical. The Ventilator Validation Kit provides ventilator manufacturers and prototype developers with a complete solution for validating new medical/hospital-grade ventilator products intended for human use. The VVK100-SYS includes […]

View All
Latest News

New Ventilator Validation Kit for Rapid Development & Field Testing

New Ventilator Validation System Allows Device Manufacturers to Test Ventilator Pressure and Volume System supports rapid development and testing of hardware for COVID-19 treatment Goleta, California – March 24, 2020 BIOPAC Systems, Inc. announced a new Ventilator Validation System for engineering and device manufacturers to use when designing and testing new and existing ventilators for […]

New Citations | VR Hazard Training for Astronauts

Space Flight Simulations Prevent Disasters Under psychological and physical stress, astronauts can put their  mission and crew in danger. Looking into this issue, researchers created a potentially catastrophic spaceflight simulation using a virtual reality environment and recorded the heart rate of participants. Psychophysiological responses to the potentially disastrous flight simulation were recorded using a BIOPAC MP […]

Read All
Request a Demonstration
Request a Demonstration