Electrodermal activity measurements

Definitions:

 
EDA = electrodermal activity
SRR = skin resistance response

SRL = skin resistance level

SCR = skin conductance response

SCL = skin conductance level

SPR = skin potential response

SPL = skin potential level

 

SRR and SCR are reciprocals of each other but are otherwise the same thing – discrete changes in skin resistance/conductance. To measure ONLY these changes, data may be acquired through a high pass filter.  The EDA100C (older model: GSR100C) has 0.05Hz and 0.5Hz HP filters which may be used for this purpose.

 

SRL and SCL are similarly reciprocols but otherwise the same thing – the baseline skin resistance/conductance. To record this form of activity, data must not be high pass filtered.  EDA100C (GSR100C) must be set up with both HP filter switches in the “DC” positions. 

 

NOTE: Recording DC signals does NOT prevent the recording of SCRs.  However, limiting the acquired data by passing it through a high pass filter for the purpose of recording ONLY SCRs prevents the recording of tonic SCL.  To take advantage of all EDA analysis features in AcqKnowledge or, more generally, to fully characterize EDA, SCL must be recorded, so HP filters should both be set to DC. 

 

While SRR, SCR, SRL and SCL rely on an external source of current (they are exosomatic), SPR and SPL do not (they are endosomatic). The EDA (GSR) series amplifiers are not appropriate for measuring SPR and SPL. (If you wish to work with these measures please contact support at BIOPAC for details.)

 

Skin conductance is usually measured in micromhos. Conductance is the reciprocal of resistance. 1 micromho of SCL is equivalent to 1,000,000 ohms SRL. The siemens and mho (as well as units derived from them such as microsiemens and micromho) are different names for the exact same unit.

 
Source: Psychophysiology Human Behavior and Physiological Response Fourth Edition
John L. Andreassi, Baruch College City University of New York,

LAWRENCE ERLBAUM ASSOCIATES, PUBLISHERS, 2000 Mahwah, New Jersey London

 

Page last modified 24Dec2014

Associated Applications

Associated Application Notes

Spotlight On
short, lightweight pinch leads

LEAD115 Lighter Lead Series

New series of unshielded 45 cm electrode leads with thin cable; suitable for facial EMG and other areas where lighter, shorter lead cables are required. Use for female Touchproof connectors to pinch clip connectors—connect electrodes to either a C-series amplifier or an MEC. The pinch connectors are light weight and the lead cable tinsel wire […]

View All
Latest News

MEAP and BIOPAC used to quantify rapid changes in cardiovascular state

The new issue of Psychophysiology has a study “Quantifying rapid changes in cardiovascular state with a moving ensemble average” using MEAP for preprocessing/analysis and BIOPAC hardware and software for data collection, display, and storage. The proof of concept study demonstrates a viable method for the adoption of ICG measures across the field of psychophysiology. MEAP (moving ensemble […]

News Citations—fMRI & Temperature, Postpartum Depression, and Fibromyalgia

BIOPAC provides software and hardware that allow for research teams to record and analyze respiration activity in physiological experimentation. Here are a few notable studies covering MRI, ECG, and Laser Doppler Flow measurement. fMRI & Thermal Perception: The neural mechanisms underlying thermal perception (how hot or cold we perceive the temperature to be) have not been fully […]

Read All
Request a Demonstration
Request a Demonstration