Electrodermal activity measurements

Definitions:

 
EDA = electrodermal activity
SRR = skin resistance response

SRL = skin resistance level

SCR = skin conductance response

SCL = skin conductance level

SPR = skin potential response

SPL = skin potential level

 

SRR and SCR are reciprocals of each other but are otherwise the same thing – discrete changes in skin resistance/conductance. To measure ONLY these changes, data may be acquired through a high pass filter.  The EDA100C (older model: GSR100C) has 0.05Hz and 0.5Hz HP filters which may be used for this purpose.

 

SRL and SCL are similarly reciprocols but otherwise the same thing – the baseline skin resistance/conductance. To record this form of activity, data must not be high pass filtered.  EDA100C (GSR100C) must be set up with both HP filter switches in the “DC” positions. 

 

NOTE: Recording DC signals does NOT prevent the recording of SCRs.  However, limiting the acquired data by passing it through a high pass filter for the purpose of recording ONLY SCRs prevents the recording of tonic SCL.  To take advantage of all EDA analysis features in AcqKnowledge or, more generally, to fully characterize EDA, SCL must be recorded, so HP filters should both be set to DC. 

 

While SRR, SCR, SRL and SCL rely on an external source of current (they are exosomatic), SPR and SPL do not (they are endosomatic). The EDA (GSR) series amplifiers are not appropriate for measuring SPR and SPL. (If you wish to work with these measures please contact support at BIOPAC for details.)

 

Skin conductance is usually measured in micromhos. Conductance is the reciprocal of resistance. 1 micromho of SCL is equivalent to 1,000,000 ohms SRL. The siemens and mho (as well as units derived from them such as microsiemens and micromho) are different names for the exact same unit.

 
Source: Psychophysiology Human Behavior and Physiological Response Fourth Edition
John L. Andreassi, Baruch College City University of New York,

LAWRENCE ERLBAUM ASSOCIATES, PUBLISHERS, 2000 Mahwah, New Jersey London

 

Page last modified 24Dec2014

Associated Applications

Associated Application Notes

Spotlight On
4-channel DAQ with AcqKnowledge and Workflow License

MP36R Systems with Workflow

This research system ships with an MP36R Data Acquisition unit and AcqKnowledge software licensed for Workflow Automation. MP36R four-channel data acquisition unit has built-in universal amplifiers that can record a wide range of physiological signals. AcqKnowledge Acquisition & Analysis Software is an interactive, intuitive program that lets you instantly view, measure, transform, and analyze data. Perform complex […]

View All
Latest News

New Citations | AcqKnowledge Research Tools

AcqKnowledge is an interactive, versatile software program that allows you to manage data in real time. Perform complex data acquisition, stimulation, triggering and analyses using intuitive navigation. Here are some recent AcqKnowledge citations… The nature of vicarious embarrassment. Journal of Business Research, 153, 355-364. Ziegler, A. H., Allen, A. M., Peloza, J., & Norris, J. I. (2022).    […]

Electrocardiography Guide Now Available

BIOPAC’s comprehensive Introductory ECG Guide addresses fundamental to advanced concerns to optimize electrocardiography data recording and analysis. Topics include: ECG Complex; Electrical and Mechanical Sequence of a Heartbeat; Systole and Diastole; Configurations for Lead I, Lead II, Lead III, 6-lead ECG, 12-lead ECG, precordial leads; Ventricular Late Potentials (VLPs); ECG Measurement Tools; Automated Analysis Routines for extracting, […]

Read All
Request a Demonstration
Request a Demonstration