Calibration values

If the Cal 1 and Cal 2 values are very close to zero, check to make sure that:
  • the amplifier is not in AC mode;
  • the filter switches (on the front of the amplifier) are not set to 0.05 Hz HP or 0.5 Hz HP;
  • the amplifier channel switch is not set to the same number as another amplifier or device (UIM100C; HLT100C)

Also, increase the gain setting if possible.

Associated Applications

Associated Application Notes

WHAT'S NEW

New Citations | BIOPAC and Mindfulness

Much of modern research aims to objectively record emotion and mindfulness in order to better...

Join the BIOPAC Community

Stay Current

Stay Connected

Request a Demonstration
Request a Demonstration