Compatibility
Pass the tubing from the chamber room through the waveguide to interface with the TSD104A Pressure Transducer in the Control room.
The output of the TSD114-MRI device is ultimately voltage; to obtain usable voltage output, the device requires amplification via BIOPAC’s DA100C. The TSD114-MRI has a signal output which is variable, and this output signal variability is linear to applied pressure in the squeeze bulb.
If the durable rubber pump bulb requires replacement, contact BIOPAC about RXPUMPBULB-MRI.
Using with third-party A/D converters
To use the TSD114-MRI with a third-party A/D converter, use DA100C and IPS100C.
DA100C amplifies and conditions the mV level signal coming from TSD114-MRIIPS100C will supply isolated power to DA100C
High level output voltage (anywhere in the range of ±10 V, such as 0-5 V) can be obtained via the front panel of IPS100C, via 3.5 mm phono plug. This signal can be directed straight to the third-party A/D converter.