Voltage Stimulators

The Gold Standard for Life Science Instruction

Stimulation for Biopac Student Lab®

Request More Info
Voltage Stimulator for BSL MP36/35
Part #: BSLSTMB, BSLSTMA

The Voltage Stimulator connects to either the MP36/MP35 or the MP30. BNC output provides compatibility with a variety of stimulator cables. The voltage Range and Level are manually controlled while the pulse timing is user-adjustable and controlled by the Biopac Student Lab software.

The Voltage Range (.025-10 Volts or .12-100 Volts) is established by a switch with a removable key. The Level is set using the knob and digital readout on the front panel. The Reference Output cable connects to any of the MP input channels for simultaneous recording of the stimulator output. A switch on the front panel allows the Reference pulse width to be “Fixed (15 ms)” so it can be recorded even when the sample rate is lower than the output pulse width.

See More...

MP356/35 Voltage Stimulator 100V
View Spec PDF

Are You Looking for These?

Part #: N/A
Categories: Stimulation - Education
Subcategories: Stimulators - Education

Alternatives
For the same Platform

Details

Compatibility


A control panel in the BSL PRO software allows for precise control of the number of pulses (select 1-254 or continuous), pulse width and pulse repetition (frequency). Pulse width and repetition can be changed “on the fly.” The stimulator can be set to start with the recording or manually at any time. Parameter limits can be established and Event markers can be activated to indicate when a change in pulse output occurs.

The stimulator can be used for the following physiological measurement protocols:

  • Twitch sub-threshold
  • Twitch threshold
  • Maximum twitch responses
  • Single twitch, summation
  • Tetanic contraction
  • Muscle tension/length versus force
  • Fatigue
  • Nerve conduction velocity

Stimulation electrodes:

  • HSTM01 Human-safe Handheld Stimulation Electrode
  • ELSTM2 Needle Electrode

Stimulation cables:

  • Nerve chamber cables: BSLCBL1A, BSLCBL2A, BSLCBL3A, and BSLCBL4B
  • Clip cables: BSLCBL7, BSLCBL11, and BSLCBL12
WARNING
Like all stimulating devices, the BSLSTM stimulator can be dangerous–please follow all stimulation safety guidelines!
  1. When powering ON the BSLSTM stimulator and/or the MP36/35 unit, follow this procedure for maximum safety: Connect the Trigger and Reference Output cables from the BSLSTM to the MP36/35 unit.
  2. Turn the MP36/35 unit ON.
  3. Launch the BSL PRO Software.
  4. Click on the MP36/35 menu and select Output Control>Stimulator.
  5. Turn the Level knob on the BSLSTM to 0 volts (fully counter-clockwise).
  6. Flip the BSLSTM Power switch ON.
  7. Attach the Output from yhe BSLSTM to the subject.
  8. Turn the Level knob to the desired initial output level.
  9. Click on the “START” button in the BSL PRO software to start acquiring data — Preferences may be set to start stimulation with start of acquisition or with the switch in the Stimulator output Control window.
  10. Incrementally adjust the Level for the desired Output.
  11. When you have completed recording and taken the required measurements, turn the Level back to 0 volts.

Support

Downloads/Resources

Knowledge Base

Recommended Items

Spotlight On
Cardiac Output amplifier

NICO Smart Amplifier

The NICO Smart Amplifier is specifically designed to measure thoracic impedance for performing noninvasive cardiac output measurements. The amplifier terminates in two connectors for impedance (Z) and its derivative (dZ), and has a 3 m cable that connects directly to an AMI100D Amplifier Input Module. Use with EL526 strip electrodes and CBL246 adapter or with EL500 […]

View All
Latest News

New Citations | BIOPAC in Exercise Physiology

Exercise physiology is the study of the body’s response to physical activity. Research in exercise physiology allows for insights in health and for a better understanding of the functions of the body at work. Here are some recent studies in exercise physiology that utilized BIOPAC tools for their research… Effect of exercise on electrocardiographic parameters […]

Electrocardiography Guide Available

BIOPAC’s comprehensive Introductory ECG Guide addresses fundamental to advanced concerns to optimize electrocardiography data recording and analysis. Topics include: ECG Complex; Electrical and Mechanical Sequence of a Heartbeat; Systole and Diastole; Configurations for Lead I, Lead II, Lead III, 6-lead ECG, 12-lead ECG, precordial leads; Ventricular Late Potentials (VLPs); ECG Measurement Tools; Automated Analysis Routines for extracting, […]

Read All
Request a Demonstration
Request a Demonstration