NIBP100C Calibration Values
Connect the NIBP100C analog output cable to the TCI105 and the DA100C amplifier. Set the DA100C amplifier to DC with a gain of 1000. In the AcqKnowledge Setup channels scaling dialog use the following values.
Cal 1 Input Voltage: 0 Volts – Map value: 0
Cal 2 Input voltage: -1 Volts – Map value: 100
Units: mmHg