BIOPAC® Systems, Inc. Logo

Multiple amplifiers per subject or multiple subjects per system

 

Shown are six independent, and synchronized, MP160 systems collecting up to 16 channels each, monitored via remote central station. Extensible to thousands of channels. #dataacquisition #physiology #lifescience #extensible #remotemonitoring #remoteneuralmonitoring #population

A post shared by BIOPAC Systems, Inc (@biopacsystems) on

SAFETY

There is no intrinsic safety problem which results from connecting multiple subjects to a single MP160/150 system. However, all such subjects are galvanically connected to each other when they are properly connected to the same MP System through wired biopotential amplifier modules. All subjects remain galvanically isolated from earth ground. (For more on this last point, see Galvanic Isolation Guidelines for the MP System.)

If there is a possibility that some such subjects may be exposed to Mains line voltage (120-240 VAC), while others may be exposed to earth ground, then the possibility exists for electrical shock hazard to all subjects connected to the single MP System. In this situation, use the IPS100C/HLT100C/OUTISO combination, in conjunction with respective subject amplifier module(s), to isolate each subject from both earth ground and from each other, or use BioNomadix modules.

PERFORMANCE

See “Grounding guidelines” for information about ground loops and how to avoid them for individual subjects. Note that if proper procedures are followed, any subject connected to a wired system for the purpose of biopotential measurements will be grounded to that system. There is no intrinsic performance problem when connecting multiple subjects to a single MP System via amplifier modules. However, since all subjects undergoing biopotential measurements are galvanically connected to the system’s ground, they are galvanically connected to each other. Consequently, if two such subjects come into direct contact with each other*, they may create a ground loop.

* Skin to skin. Most clothing is made of electrically insulating material, so two subjects touching through clothing will not present the same concerns as two subjects touching directly.

If EDA measurements are being performed on multiple subjects, all of whom are connected to the same MP System using EDA100/100C or GSR100/100C, and if the subjects touch each other, galvanic currents will flow between the subjects, thus disturbing the measurement. The same situation holds true with the EBI100C or NICO100C amplifiers. If this situation is not acceptable, use the IPS100C/HLT100C/OUTISO combination to isolate the amplifiers from one another. Alternatively, data may be acquired with 100D series Smart Ampliers or wireless BioNomadix modules. The BioNomadix transmitters are fully isolated thus there will be no interaction possible between the EDA and Impedance Cardiography on the same subject or between subjects.

There is no intrinsic performance problem which results when conducting multiple biopotential measurements on the same subject. In this case, it’s only important to have simply ONE ground lead attached to the subject. There are some special cases to consider for certain amplifier modules and combinations of certain modules.

Special Cases

See “Grounding guidelines” for information about the use of EDA100C (GSR100C) and other biopotential amplifiers.

When performing multiple EDA (GSR) measurements on the same subject

All EDA100C (GSR100C) amplifiers should share the same ground connection point on the subject. That is, only one of these amplifiers should have a lead wire connecting VIn- to the subject. The current from each of the EDA amplifiers will return through this one site. If this situation is not acceptable, then each additional EDA100C amplifier should be galvanically isolated from the others using the IPS100C/HLT100C/OUTISO combination, or an EDA100D Smart Amplifier or BN-PPGED wireless module should be used in place of EDA100C or GSR100C.

When using the EBI100C or NICO100C amplifier

Use the EBI100C or NICO100C ground lead (the I IN connection) as the ground for all attached amplifiers. As an alternative, it’s always possible to galvanically isolate the EBI100C or NICO100C from the other attached amplifiers by connecting through  IPS100C/HLT100C/OUTISO/CBL100. Alternatively, the BN-NICO may be used instead of the EBI100C (as long as phase information is not needed) or NICO100C.

When using multiple EBI100C or NICO100C amplifiers on the same subject

All EBI100C or NICO100C amplifiers should share the same ground connection point on the subject. That is, only one module should have a lead wire connecting I IN to the subject. The current return path for all bioimpedance amplifiers will return through this one site. If this situation is not acceptable, then each EBI100C or NICO100C amplifier should be galvanically isolated from the others using the IPS100C/HLT100C/OUTISO combination. Again, BN-NICO modules may be used to replace one or all of these modules (as long as phase information is not required).

When using the EDA/GSR100C in combination with an EBI100C or NICO100C

Use the EBI100C or NICO current return (I IN) to ground the EDA100C too. In this case, the EDA100C will effectively be measuring conductance between the chosen fingertip and the ground position for the EBI100C or NICO100C (typically the abdomen). For relative EDA (GSR) changes, this configuration is functional. However, for fingertip-to-fingertip EDA (GSR) measurements, while also performing EBI100C or NICO100C measurements, the EDA100C should be galvanically isolated from the other amplifiers by using the IPS100C/HLT100C/OUTISO combination. Alternatively, EDA100D, BN-NICO or BN-PPGED may be substituted for the corresponding 100C modules.

Associated Applications

Associated Application Notes

Spotlight On
Ventilator Validation Kit

Ventilator Validation and Test Equipment

The Ventilator Validation Kit supports new prototype development, existing design validation for production testing, mobile field testing and verification, and other uses where accurate pressure and flow measurements are critical. The Ventilator Validation Kit provides ventilator manufacturers and prototype developers with a complete solution for validating new medical/hospital-grade ventilator products intended for human use. The VVK100-SYS includes […]

View All
Latest News

New Ventilator Validation Kit for Rapid Development & Field Testing

New Ventilator Validation System Allows Device Manufacturers to Test Ventilator Pressure and Volume System supports rapid development and testing of hardware for COVID-19 treatment Goleta, California – March 24, 2020 BIOPAC Systems, Inc. announced a new Ventilator Validation System for engineering and device manufacturers to use when designing and testing new and existing ventilators for […]

New Citations | VR Hazard Training for Astronauts

Space Flight Simulations Prevent Disasters Under psychological and physical stress, astronauts can put their  mission and crew in danger. Looking into this issue, researchers created a potentially catastrophic spaceflight simulation using a virtual reality environment and recorded the heart rate of participants. Psychophysiological responses to the potentially disastrous flight simulation were recorded using a BIOPAC MP […]

Read All
Request a Demonstration
Request a Demonstration