There is no intrinsic safety problem which results from connecting multiple subjects to a single MP160/150 system. However, all such subjects are galvanically connected to each other when they are properly connected to the same MP System through wired biopotential amplifier modules. All subjects remain galvanically isolated from earth ground. (For more on this last point, see Galvanic Isolation Guidelines for the MP System.)
If there is a possibility that some such subjects may be exposed to Mains line voltage (120-240 VAC), while others may be exposed to earth ground, then the possibility exists for electrical shock hazard to all subjects connected to the single MP System. In this situation, use the IPS100C/HLT100C/OUTISO combination, in conjunction with respective subject amplifier module(s), to isolate each subject from both earth ground and from each other, or use BioNomadix modules.
See “Grounding guidelines” for information about ground loops and how to avoid them for individual subjects. Note that if proper procedures are followed, any subject connected to a wired system for the purpose of biopotential measurements will be grounded to that system. There is no intrinsic performance problem when connecting multiple subjects to a single MP System via amplifier modules. However, since all subjects undergoing biopotential measurements are galvanically connected to the system’s ground, they are galvanically connected to each other. Consequently, if two such subjects come into direct contact with each other*, they may create a ground loop.
* Skin to skin. Most clothing is made of electrically insulating material, so two subjects touching through clothing will not present the same concerns as two subjects touching directly.
If EDA measurements are being performed on multiple subjects, all of whom are connected to the same MP System using EDA100/100C or GSR100/100C, and if the subjects touch each other, galvanic currents will flow between the subjects, thus disturbing the measurement. The same situation holds true with the EBI100C or NICO100C amplifiers. If this situation is not acceptable, use the IPS100C/HLT100C/OUTISO combination to isolate the amplifiers from one another. Alternatively, data may be acquired with 100D series Smart Ampliers or wireless BioNomadix modules. The BioNomadix transmitters are fully isolated thus there will be no interaction possible between the EDA and Impedance Cardiography on the same subject or between subjects.
There is no intrinsic performance problem which results when conducting multiple biopotential measurements on the same subject. In this case, it’s only important to have simply ONE ground lead attached to the subject. There are some special cases to consider for certain amplifier modules and combinations of certain modules.
See “Grounding guidelines” for information about the use of EDA100C (GSR100C) and other biopotential amplifiers.
When performing multiple EDA (GSR) measurements on the same subject
All EDA100C (GSR100C) amplifiers should share the same ground connection point on the subject. That is, only one of these amplifiers should have a lead wire connecting VIn- to the subject. The current from each of the EDA amplifiers will return through this one site. If this situation is not acceptable, then each additional EDA100C amplifier should be galvanically isolated from the others using the IPS100C/HLT100C/OUTISO combination, or an EDA100D Smart Amplifier or BN-PPGED wireless module should be used in place of EDA100C or GSR100C.
When using the EBI100C or NICO100C amplifier
Use the EBI100C or NICO100C ground lead (the I IN connection) as the ground for all attached amplifiers. As an alternative, it’s always possible to galvanically isolate the EBI100C or NICO100C from the other attached amplifiers by connecting through IPS100C/HLT100C/OUTISO/CBL100. Alternatively, the BN-NICO may be used instead of the EBI100C (as long as phase information is not needed) or NICO100C.
When using multiple EBI100C or NICO100C* amplifiers on the same subject
All EBI100C or NICO100C amplifiers should share the same ground connection point on the subject. That is, only one module should have a lead wire connecting I IN to the subject. The current return path for all bioimpedance amplifiers will return through this one site. If this situation is not acceptable, then each EBI100C or NICO100C amplifier should be galvanically isolated from the others using the IPS100C/HLT100C/OUTISO combination. Again, BN-NICO modules may be used to replace one or all of these modules (as long as phase information is not required). *Important: Do no use two NICO100D Smart Amplifiers on the same subject!
When using the EDA/GSR100C in combination with an EBI100C or NICO100C
Use the EBI100C or NICO current return (I IN) to ground the EDA100C too. In this case, the EDA100C will effectively be measuring conductance between the chosen fingertip and the ground position for the EBI100C or NICO100C (typically the abdomen). For relative EDA (GSR) changes, this configuration is functional. However, for fingertip-to-fingertip EDA (GSR) measurements, while also performing EBI100C or NICO100C measurements, the EDA100C should be galvanically isolated from the other amplifiers by using the IPS100C/HLT100C/OUTISO combination. Alternatively, EDA100D, BN-NICO or BN-PPGED may be substituted for the corresponding 100C modules.
Workflow—A Great Tool for Every Researcher! Automate AcqKnowledge with New Workflow Tool New automation tool significantly saves time and improves consistency with standardized procedures Workflow’s drag-and-drop interface allows you to automate and repeat the signal-conditioning and processing steps your protocol requires in an easy-to-use process without programming or scripting knowledge. Batch process multiple files with confidence […]
View AllExercise physiology is the study of the body’s response to physical activity. Research in exercise physiology allows for insights in health and for a better understanding of the functions of the body at work. Here are some recent studies in exercise physiology that utilized BIOPAC tools for their research… Effect of exercise on electrocardiographic parameters […]
BIOPAC’s comprehensive Introductory ECG Guide addresses fundamental to advanced concerns to optimize electrocardiography data recording and analysis. Topics include: ECG Complex; Electrical and Mechanical Sequence of a Heartbeat; Systole and Diastole; Configurations for Lead I, Lead II, Lead III, 6-lead ECG, 12-lead ECG, precordial leads; Ventricular Late Potentials (VLPs); ECG Measurement Tools; Automated Analysis Routines for extracting, […]
Read All