BIOPAC® Systems, Inc. Logo

130 – Blood Pressure Measurement for the BSL PRO

The most common form of indirect blood pressure measurement employs a pressure cuff, pump and pressure transducer. This complete assembly is commonly referred to as a Sphygmomanometer.

Typically, the cuff is wrapped around the upper arm and is inflated to a pressure exceeding that of the brachial artery. This amount of pressure collapses the artery and stops the flow of blood to the arm. The pressure of the cuff is slowly reduced as the pressure in the cuff is monitored by the pressure transducer. As the pressure drops, it will eventually match the systolic (peak) arterial pressure. At this point, the blood is able to “squirt” through the brachial artery. This squirting results in turbulence which creates the Korotkoff sounds. The Korotkoff sounds are detected using an SS17L physiological sounds transducer. As the cuff pressure continues to drop, the pressure will eventually match the diastolic pressure of the artery. At this point the Korotkoff sounds stop completely, because the blood is now flowing unrestricted through the artery.

Associated Applications

Spotlight On
Ventilator Validation Kit

Ventilator Validation and Test Equipment

The Ventilator Validation Kit supports new prototype development, existing design validation for production testing, mobile field testing and verification, and other uses where accurate pressure and flow measurements are critical. The Ventilator Validation Kit provides ventilator manufacturers and prototype developers with a complete solution for validating new medical/hospital-grade ventilator products intended for human use. The VVK100-SYS includes […]

View All
Latest News

Researchers Respond | iSAVE: Rapidly deployable system to expand ventilator capacity

MIT and Harvard researchers developed a new technique to expand ventilator capacity Individualized System for Augmenting Ventilator Efficacy (iSAVE): A Rapidly deployable system to expand ventilator capacity “The iSAVE is a rapidly deployable solution in response to the urgent shortage of mechanical ventilators for respiratory support during the COVID-19 pandemic. In contrast to previously described […]

New Ventilator Validation Kit for Rapid Development & Field Testing

New Ventilator Validation System Allows Device Manufacturers to Test Ventilator Pressure and Volume System supports rapid development and testing of hardware for COVID-19 treatment Goleta, California – March 24, 2020 BIOPAC Systems, Inc. announced a new Ventilator Validation System for engineering and device manufacturers to use when designing and testing new and existing ventilators for […]

Read All
Request a Demonstration
Request a Demonstration