BSL PRO optimize range

The BSL MP30/35/36 unit has four isolated, human-safe universal biopotential/transducer amplifiers that are controlled by the BSLPRO software. When recording data, it is important to set the gain of the amplifier as high as possible without clipping it. If you are unsure what gain to use for your signal type, the Optimize Ranges feature will adjust the screen display so that you can see the incoming signal relative to the range of the A/D converter. The Optimize Ranges option is located from the Display menu, by right mouse clicking, and also as a toolbar icon . A green bar located at the right edge of the channel display identifies the vertical range.

If the signal is very small, relative to the display window, increase the gain so that the signal fills more of the channel display region. This will maximize the resolution of the signal relative to the input signal range of the A/D converter. The following screen shots show an ECG signal with the gain set to 1K, 5K, and 10K. By increasing the gain of the amplifier to 10K, the signal is using the full range of the A/D converter.

Note: The Optimize Ranges feature only works on analog input channels.
 
The following screen shows data with the default ECG gain setting (1K).
 
The following screen shows the data with the gain set to 5K.
 
The following screen shows the data with the gain set to 10K.
 
The gain is adjusted in the Setup channels > View change parameters dialog

Associated Disciplines

Spotlight On
Ventilator Validation Kit

Ventilator Validation and Test Equipment

The Ventilator Validation Kit supports new prototype development, existing design validation for production testing, mobile field testing and verification, and other uses where accurate pressure and flow measurements are critical. The Ventilator Validation Kit provides ventilator manufacturers and prototype developers with a complete solution for validating new medical/hospital-grade ventilator products intended for human use. The VVK100-SYS includes […]

View All
Latest News

Researchers Respond | iSAVE: Rapidly deployable system to expand ventilator capacity

MIT and Harvard researchers developed a new technique to expand ventilator capacity Individualized System for Augmenting Ventilator Efficacy (iSAVE): A Rapidly deployable system to expand ventilator capacity “The iSAVE is a rapidly deployable solution in response to the urgent shortage of mechanical ventilators for respiratory support during the COVID-19 pandemic. In contrast to previously described […]

New Ventilator Validation Kit for Rapid Development & Field Testing

New Ventilator Validation System Allows Device Manufacturers to Test Ventilator Pressure and Volume System supports rapid development and testing of hardware for COVID-19 treatment Goleta, California – March 24, 2020 BIOPAC Systems, Inc. announced a new Ventilator Validation System for engineering and device manufacturers to use when designing and testing new and existing ventilators for […]

Read All
Request a Demonstration
Request a Demonstration