BIOPAC® Systems, Inc. Logo

Amplifier module filter response times

The MP series amplifier modules incorporate a variety of filtering options. The low pass filtering options have an effect on the signal response time, which is sometimes referred to as signal delay.

Signal delay is often plotted as the function “group delay” versus frequency. This type of plot shows the typical delay the filter will have for a wide range of frequencies. The group delay plot is the derivative of the filter phase plot with respect to frequency. If the filter is perfectly linear phase, the group delay plot will be a straight horizontal line, because the derivative of a constant (linear) slope is a constant.

In practice it’s often difficult to utilize a group delay plot to get a quick and simple handle on essential filter signal delay, unless one is experienced in reading such plots. Instead, it’s typically better to show the filter response to a well-understood input signal, such as a step function.

The following plots illustrate the delay times for a variety of low pass filter settings.

10 Hz Low Pass Filter – 4 pole Besselworth
10 Hz Low Pass Filter – 4 pole Besselworth

35 Hz Low Pass Filter – 4 pole Besselworth
35 Hz Low Pass Notch Filter – 4 pole Besselworth – Notch at 60 Hz

100 Hz Low Pass Filter – 4 pole Besselworth
100 Hz Low Pass Filter – 4 pole Besselworth

500 Hz Low Pass Filter – 4 pole Besselworth
500 Hz Low Pass Filter – 4 pole Besselworth

Note that signal delay is proportional to the cut-off frequency for any particular 4 pole low pass Besselworth filter. The one minor exception is the 35 Hz LPN filter, because it consists of a 4 pole Besselworth filter and a 60 Hz Notch (band reject) filter. This additional filter adds a small additional delay.

Step Response Signal Delay (approximate)

Filter Type:

10 Hz LP

35 Hz LP

100 Hz LP

500 Hz LP

Delay at 50% (approximate):

30 ms

11 ms

3 ms

0.6 ms

Spotlight On
Ventilator Validation Kit

Ventilator Validation and Test Equipment

The Ventilator Validation Kit supports new prototype development, existing design validation for production testing, mobile field testing and verification, and other uses where accurate pressure and flow measurements are critical. The Ventilator Validation Kit provides ventilator manufacturers and prototype developers with a complete solution for validating new medical/hospital-grade ventilator products intended for human use. The VVK100-SYS includes […]

View All
Latest News

Researchers Respond | iSAVE: Rapidly deployable system to expand ventilator capacity

MIT and Harvard researchers developed a new technique to expand ventilator capacity Individualized System for Augmenting Ventilator Efficacy (iSAVE): A Rapidly deployable system to expand ventilator capacity “The iSAVE is a rapidly deployable solution in response to the urgent shortage of mechanical ventilators for respiratory support during the COVID-19 pandemic. In contrast to previously described […]

New Ventilator Validation Kit for Rapid Development & Field Testing

New Ventilator Validation System Allows Device Manufacturers to Test Ventilator Pressure and Volume System supports rapid development and testing of hardware for COVID-19 treatment Goleta, California – March 24, 2020 BIOPAC Systems, Inc. announced a new Ventilator Validation System for engineering and device manufacturers to use when designing and testing new and existing ventilators for […]

Read All
Request a Demonstration
Request a Demonstration