Measure Impedance Magnitude (Zo) and Derivative of Impedance Magnitude (dZ/dt)

Noninvasive bioimpedance technique to measure relative differences in Cardiac Output and Stroke Volume

Request More Info
View Cart “Abrasive Pads 10/pk” has been added to your cart.
Noninvasive Cardiac Output Sensor
Part #: SS31LA

The SS31LA simultaneously measures Impedance Magnitude (Zo; labeled “Z” on the module) and Derivative of Impedance Magnitude (dZ/dt; labeled “DZ” on the module). The SS31LA noninvasively records the parameters associated with Cardiac Output measurements.

  • Measure changes in Cardiac Output under a variety of conditions: laying down, sitting up, standing up, and post-exercise.
  • Use the Kubicek equation to determine Stroke Volume and Cardiac Output.
  • For respiration measurements, observe the associated variation in the Impedance Magnitude “Z” signal.

Use with an 8 spot electrode lead configuration and use on stationary subjects—the SS31LA is sensitive to motion artifact.

See More...

NICO transducer
View Spec PDF

Are You Looking for These?

Part #: SS31LA
Categories: Transducers - Education, Transducers - Research
Subcategories: Cardiac Output - Education, Cardiac Output - Research

Equivalents
With a different Platform

Details

Compatibility


The SS31LA incorporates a precision high-frequency current source, which injects a small (400 µA rms @ 100 kHz sine wave) current through the measurement tissue volume defined by the placement of a set of current source electrodes. A separate set of monitoring electrodes then measures the voltage developed across the tissue volume. Because the current is constant, the voltage measured is proportional to the characteristics of the biological impedance of the tissue volume.

See BSL PRO Lesson H21 Impedance Cardiography for sample setup and data, as well as an extensively updated physiological introduction that covers Cardiac Output as a Function of Stroke Volume and Heart Rate, Cardiac Output as Related to Mean Arterial Pressure and Peripheral Resistance, How to Determine Cardiac Output, Principles of Impedance Cardiography (ICG), and Ideas for eliciting cardiovascular responses through physical maneuvers and psychological tasks.


Note: SS31LA replaces the SS31L, which had lead connectors designed for strip electrodes, such as EL506, which were discontinued due to manufacturing limitations.

Support

Downloads/Resources

Knowledge Base

Recommended Items

Spotlight On
BSL Practical Lab Kit

Biopac Student Lab Human Physiology Systems

BSL System with MP36 and hardware for 11 guided lessons plus 14 dynamic lessons and student-designed Human Physiology experiments. Use the power of the BSL Human Physiology Teaching System to conduct the most widely studied responses in physiology teaching labs. Lessons target the circulatory system, respiratory system, muscular function, brain function, ANS, exercise physiology and […]

View All
Latest News

New Citations | BIOPAC’s AcqKnowledge Software

BIOPAC tools are used for studies in a variety of different fields and disciplines. AcqKnowledge software is one of our universally applicable tools, used by thousands of labs around the world and cited in thousands of peer reviewed publications, such as the following studies… The psychophysiology of guilt in healthy adults. Cognitive, Affective, & Behavioral Neuroscience, 1-18. […]

New Citations | BIOPAC in Human Physiology and 4T|Phys Recap

We had the opportunity to meet lots of researchers at our 4T|Phys Human Physiology Conference this month and demonstrate solutions for life science studies. We presented 39 unique skills sessions, including sessions on measuring blood pressure, recording HRV, and utilizing VR. These studies all apply these skills and further demonstrate how to apply our equipment… Real-Time Personalized Physiologically […]

Read All
Request a Demonstration
Request a Demonstration