Calibration Kit for SS11LB

Request More Info
SS11LB validation and calibration template
Part #: FLOWCAL

Optional calibration/validation kit

Use FLOWCAL to validate factory calibration of the SS11LB. This kit connects the syringe to the SS11LB, coupler, and template file with calibration guide.

  • The SS11LB is factory-calibrated for use when directly connected to a mouthpiece. If the flow transducer is connected to a hose, facemask, or other tubing it should be recalibrated with those attachments by using the syringe and this kit.
  • Users wishing to perform an accurate validation should also have equipment that can measure humidity, temperature, and pressure of the lab environment.

Download the free template file and guide below.

Shipping now: Calibrated 3 L calibration syringe (AFT27) plus coupler (AFT11D)

Older version: Calibrated 2 L calibration syringe (AFT26) plus coupler (AFT11D)

See More...

Are You Looking for These?

Part #: FLOWCAL
Categories: Gas Analysis - Education, Airflow & Gas Analysis - Research
Subcategories: Calibration - Education, Calibration - Research

Details

Compatibility


Optional SS11LB Calibration

Use this to perform a user calibration of an SS11LB airflow transducer. An MP unit, SS11LB, and FLOWCAL (AFT27/26 syringe + AFT11D coupler) are required. The SS11LB must be warmed up for at least 10 minutes prior to calibration.

User calibration settings are stored within each SS11LB and are used automatically by the SS11LB analog channel preset, BSL Lessons, and BSL PRO Lessons.

  1. Download and save the graph template for the calibration syringe you will use

3 L Syringe – FLOWCAL graph template for AFT27

2 L Syringe – FLOWCAL graph template for AFT26

  1. Connect the SS11LB to CH1. Keep the SS11LB connected to the MP unit and the MP unit powered on during a minimum 10-minute warm-up period.
  2. Launch AcqKnowledge or Biopac Student Lab.
  3. Open the graph template file.
  4. Click the approporiate button(s) for the deisred action:
  • Calibrate the SS11LB: click”Calibrate” and follow the user calibration procedure outlined in the FLOWCAL Guide.
  • Remove any user calibration and return to the factory calibration values: click “Restore Factory Calibration.”
  • See the current calibration settings stored in the SS11LB: click “Info.”

Support

Knowledge Base

Spotlight On
B-Alert EEG+ECG and AcqKnowledge

B-Alert Wireless EEG System

Easily Add EEG to Your Research Study! B-Alert X-Series Systems offer acquisition of high fidelity wireless EEG plus ECG, and AcqKnowledge® adds powerful analysis tools, including automated scoring and reporting options. Choose X10 for 9-channels EEG+1 biopotential or X24 for 20-channel EEG+1 biopotential. Together, a B-Alert X-Series System and AcqKnowledge provide researchers with a turnkey, easily procured system solution […]

View All
Latest News

New Citations | How Sudoku Helps With Rehabilitation & More!

Rehabilitation with Sudoku Sudoku, in the last 10 years, has seen a resurgence in popularity as a challenging logic game. Scientists have also fallen in love with the math puzzle, as a useful tool for neurorehabilitation and cognitive workouts. This study measured prefrontal cortical activity with a continuous-wave fNIR Imager to observe how participants solved […]

Electrocardiography Guide Now Available

BIOPAC’s comprehensive Introductory ECG Guide addresses fundamental to advanced concerns to optimize electrocardiography data recording and analysis. Topics include: ECG Complex; Electrical and Mechanical Sequence of a Heartbeat; Systole and Diastole; Configurations for Lead I, Lead II, Lead III, 6-lead ECG, 12-lead ECG, precordial leads; Ventricular Late Potentials (VLPs); ECG Measurement Tools; Automated Analysis Routines for extracting, […]

Read All
Request a Demonstration
Request a Demonstration