BIOPAC® Systems, Inc. Logo

Calibration Kit for SS11LB

Request More Info
SS11LB validation and calibration template
Part #: FLOWCAL

Optional calibration/validation kit

  • Validate factory calibration of the SS11LB
  • Calibrate TSD117A/TSD117A-MRI for use with SS11LB

This kit connects the syringe to the SS11LB/TSD117A/TSD117A-MRI. Download the free template file and guide below.

Shipping now: Calibrated 3 L calibration syringe (AFT27) plus coupler (AFT11D)

Older version: Calibrated 2 L calibration syringe (AFT26) plus coupler (AFT11D)

Users wishing to perform an accurate validation should also have equipment that can measure humidity, temperature, and pressure of the lab environment.

The SS11LB is factory-calibrated for use when directly connected to a mouthpiece. If the flow transducer is connected to a hose, facemask, or other tubing it should be recalibrated with those attachments by using the syringe and this kit.

See More...

Are You Looking for These?

Part #: FLOWCAL
Categories: Airflow & Gas Analysis - Research, Gas Analysis - Education
Subcategories: Calibration - Research, Calibration - Education

Details

Compatibility


Optional SS11LB Calibration

Use this to perform a user calibration of an SS11LB airflow transducer. An MP unit, SS11LB, and FLOWCAL (AFT27/26 syringe + AFT11D coupler) are required. The SS11LB must be warmed up for at least 10 minutes prior to calibration.

User calibration settings are stored within each SS11LB and are used automatically by the SS11LB analog channel preset, BSL Lessons, and BSL PRO Lessons.

  1. Download and save the graph template for the calibration syringe you will use

3 L Syringe – FLOWCAL graph template for AFT27

2 L Syringe – FLOWCAL graph template for AFT26

  1. Connect the SS11LB to CH1. Keep the SS11LB connected to the MP unit and the MP unit powered on during a minimum 10-minute warm-up period.
  2. Launch AcqKnowledge or Biopac Student Lab.
  3. Open the graph template file.
  4. Click the approporiate button(s) for the deisred action:
  • Calibrate the SS11LB: click”Calibrate” and follow the user calibration procedure outlined in the FLOWCAL Guide.
  • Remove any user calibration and return to the factory calibration values: click “Restore Factory Calibration.”
  • See the current calibration settings stored in the SS11LB: click “Info.”

Support

Knowledge Base

Spotlight On
Smart Center controller, 2-3 wireless transmitters & AcqKnowledge plus case

Smart Center | Stand-alone BioNomadix Wireless Systems

Smart Center Stand-alone, Portable Wireless Physiology Systems Smart Center Essentials entry-level system includes a Smart Center Device (BN-SMART with USB power cable), 2 or 3 Transmitters (BN-xxx-T with chargers); AcqKnowledge for Smart Center, and a case. Smart Center Enhanced adds a Logger (BN-LOGGER) and Basic Scripting License (ACK100W-BAS) to the Smart Center Device (BN-SMART with powercable), 3 […]

View All
Latest News

Students Get Smart

Science Night at Adams Elementary (Santa Barbara, CA) encourages students to discover, play, explore and learn about the importance of science and engineering in daily life. BIOPAC set up a BioNomadix Wireless Smart Center to allow students to record their heart rate. Electrodes were placed on students and leads were connected to a wireless transmitter that sent data […]

New Citations | Stimulation and Mental Processing

Recent developments regarding electric stimulation and mental processing have lead to interesting developments in psychophysiology and physiological sciences. The following studies look at developments in how we potentially care for and understand solutions to issues with mental health, as well as new ways to monitor electrical stimulus on participants. The ventromedial prefrontal cortex (vmPFC) is […]

Read All
Request a Demonstration
Request a Demonstration