SS11LB validation and calibration template
Part #: FLOWCAL

Optional calibration/validation kit

  • Validate factory calibration of the SS11LB
  • Calibrate TSD117A/TSD117A-MRI for use with SS11LB

This kit connects the syringe to the SS11LB/TSD117A/TSD117A-MRI. Download the free template file and guide below.

Shipping now: Calibrated 3 L calibration syringe (AFT27) plus coupler (AFT11D)

Older version: Calibrated 2 L calibration syringe (AFT26) plus coupler (AFT11D)

Users wishing to perform an accurate validation should also have equipment that can measure humidity, temperature, and pressure of the lab environment.

The SS11LB is factory-calibrated for use when directly connected to a mouthpiece. If the flow transducer is connected to a hose, facemask, or other tubing it should be recalibrated with those attachments by using the syringe and this kit.

See More...

Are You Looking for These?

Part #: FLOWCAL
Categories: Gas Analysis - Education, Airflow & Gas Analysis - Research
Subcategories: Calibration - Education, Calibration - Research

Details

Compatibility


Optional SS11LB Calibration

Use this to perform a user calibration of an SS11LB airflow transducer. An MP unit, SS11LB, and FLOWCAL (AFT27/26 syringe + AFT11D coupler) are required. The SS11LB must be warmed up for at least 10 minutes prior to calibration.

User calibration settings are stored within each SS11LB and are used automatically by the SS11LB analog channel preset, BSL Lessons, and BSL PRO Lessons.

  1. Download and save the graph template for the calibration syringe you will use

3 L Syringe – FLOWCAL graph template for AFT27

2 L Syringe – FLOWCAL graph template for AFT26

  1. Connect the SS11LB to CH1. Keep the SS11LB connected to the MP unit and the MP unit powered on during a minimum 10-minute warm-up period.
  2. Launch AcqKnowledge or Biopac Student Lab.
  3. Open the graph template file.
  4. Click the approporiate button(s) for the deisred action:
  • Calibrate the SS11LB: click”Calibrate” and follow the user calibration procedure outlined in the FLOWCAL Guide.
  • Remove any user calibration and return to the factory calibration values: click “Restore Factory Calibration.”
  • See the current calibration settings stored in the SS11LB: click “Info.”

Support

Knowledge Base

Spotlight On
wireless EEG ECG telemetry for rodents

Epoch Wireless ECG Systems

Long-term wireless ECG system Wireless ECG implantable transmitters for mice or rats Additional transmitters available—just move the receiver tray to record from multiple home cages The Epoch ECG Transmitter amplifies and transmits high-fidelity ECG data. Implants are small enough to be implanted into mice as young as P18. To record, simply place the animal’s cage on top of the […]

View All
Latest News

New Citations | Psychophysiology, Chewing Study, Pulse Wave Velocity

Recent studies in Psychophysiology, EMG, ECG and PPG measurement utilizing BIOPAC software and hardware: Fight or Flight (Reflex Response Time) “Fight or flight” is a physiological response that occurs due to a perceived threat to survival. It can physiologically be observed through an increase in heart rate, respiratory rate, blood pressure, oxygen availability, and other mechanisms. In […]

New Citations | Heart Rate & Heart Rate Variability, Wingate Anaerobic Test, and Neuroscience of Music

HRV, EMG, and EEG studies using BIOPAC hardware and software include: DON’T STRESS Humans encounter various stressful situations everyday at work, home, and school. Such stress when experienced at high degrees and/or for a long duration of time could lead to cardiovascular diseases, cognitive dysfunctions, and psychological disorders. This study examined the validity of a novel physiological measurement […]

Read All
Request a Demonstration
Request a Demonstration