Calibration Kit for SS11LB

Request More Info
SS11LB validation and calibration template
Part #: FLOWCAL

Optional calibration/validation kit

  • Validate factory calibration of the SS11LB
  • Calibrate TSD117A/TSD117A-MRI for use with SS11LB

This kit connects the syringe to the SS11LB/TSD117A/TSD117A-MRI. Download the free template file and guide below.

Shipping now: Calibrated 3 L calibration syringe (AFT27) plus coupler (AFT11D)

Older version: Calibrated 2 L calibration syringe (AFT26) plus coupler (AFT11D)

Users wishing to perform an accurate validation should also have equipment that can measure humidity, temperature, and pressure of the lab environment.

The SS11LB is factory-calibrated for use when directly connected to a mouthpiece. If the flow transducer is connected to a hose, facemask, or other tubing it should be recalibrated with those attachments by using the syringe and this kit.

See More...

Are You Looking for These?

Part #: FLOWCAL
Categories: Gas Analysis - Education, Airflow & Gas Analysis - Research
Subcategories: Calibration - Education, Calibration - Research

Details

Compatibility


Optional SS11LB Calibration

Use this to perform a user calibration of an SS11LB airflow transducer. An MP unit, SS11LB, and FLOWCAL (AFT27/26 syringe + AFT11D coupler) are required. The SS11LB must be warmed up for at least 10 minutes prior to calibration.

User calibration settings are stored within each SS11LB and are used automatically by the SS11LB analog channel preset, BSL Lessons, and BSL PRO Lessons.

  1. Download and save the graph template for the calibration syringe you will use

3 L Syringe – FLOWCAL graph template for AFT27

2 L Syringe – FLOWCAL graph template for AFT26

  1. Connect the SS11LB to CH1. Keep the SS11LB connected to the MP unit and the MP unit powered on during a minimum 10-minute warm-up period.
  2. Launch AcqKnowledge or Biopac Student Lab.
  3. Open the graph template file.
  4. Click the approporiate button(s) for the deisred action:
  • Calibrate the SS11LB: click”Calibrate” and follow the user calibration procedure outlined in the FLOWCAL Guide.
  • Remove any user calibration and return to the factory calibration values: click “Restore Factory Calibration.”
  • See the current calibration settings stored in the SS11LB: click “Info.”

Support

Knowledge Base

Spotlight On
continuous NIR spectroscopy (NIRS) module

fNIR Imaging Systems

fNIR Imaging Systems are stand-alone functional brain imaging solutions for continuous NIR spectroscopy (NIRS). They provide an in-lab cognitive function assessment and eliminate a great many of the drawbacks of a functional MRI. fNIR optical imaging technology measures neural activity and hemodynamic response in the prefrontal cortex. Subjects wear an fNIR sensor (IR light sources […]

View All
Latest News

New Citations | Research in Motion

Physiological researchers have been studying the lower body through examining impacts of training, movement instruction, and innovative research measurements on an individual’s well being. The following articles represent recent developments in physiological research, moving our understanding of lower limbs and the body as a whole forward, one step at a time. The Importance of Form Improper […]

New Citations | Computer-Human Interaction

Technology is more interactive in a multitude of contexts, from completing tasks at work, looking for help online, to leisurely activities like news and video streaming. Understanding how to improve interaction between humans and computers can hold benefits in multiple situations. The following are recent human-computer interaction (HCI) studies. The Eyes Have It Nasser et al. […]

Read All
Request a Demonstration
Request a Demonstration