Calibration Kit for SS11LB

Request More Info
SS11LB validation and calibration template
Part #: FLOWCAL

Optional calibration/validation kit

  • Validate factory calibration of the SS11LB
  • Calibrate TSD117A/TSD117A-MRI for use with SS11LB

This kit connects the syringe to the SS11LB/TSD117A/TSD117A-MRI. Download the free template file and guide below.

Shipping now: Calibrated 3 L calibration syringe (AFT27) plus coupler (AFT11D)

Older version: Calibrated 2 L calibration syringe (AFT26) plus coupler (AFT11D)

Users wishing to perform an accurate validation should also have equipment that can measure humidity, temperature, and pressure of the lab environment.

The SS11LB is factory-calibrated for use when directly connected to a mouthpiece. If the flow transducer is connected to a hose, facemask, or other tubing it should be recalibrated with those attachments by using the syringe and this kit.

See More...

Are You Looking for These?

Part #: FLOWCAL
Categories: Gas Analysis - Education, Airflow & Gas Analysis - Research
Subcategories: Calibration - Education, Calibration - Research

Details

Compatibility


Optional SS11LB Calibration

Use this to perform a user calibration of an SS11LB airflow transducer. An MP unit, SS11LB, and FLOWCAL (AFT27/26 syringe + AFT11D coupler) are required. The SS11LB must be warmed up for at least 10 minutes prior to calibration.

User calibration settings are stored within each SS11LB and are used automatically by the SS11LB analog channel preset, BSL Lessons, and BSL PRO Lessons.

  1. Download and save the graph template for the calibration syringe you will use

3 L Syringe – FLOWCAL graph template for AFT27

2 L Syringe – FLOWCAL graph template for AFT26

  1. Connect the SS11LB to CH1. Keep the SS11LB connected to the MP unit and the MP unit powered on during a minimum 10-minute warm-up period.
  2. Launch AcqKnowledge or Biopac Student Lab.
  3. Open the graph template file.
  4. Click the approporiate button(s) for the deisred action:
  • Calibrate the SS11LB: click”Calibrate” and follow the user calibration procedure outlined in the FLOWCAL Guide.
  • Remove any user calibration and return to the factory calibration values: click “Restore Factory Calibration.”
  • See the current calibration settings stored in the SS11LB: click “Info.”

Support

Knowledge Base

Spotlight On
fNIR Education System Imager

fNIR Imaging Education System

fNIR Imaging Systems are stand-alone functional brain imaging solutions for continuous NIR spectroscopy (NIRS). They provide an in-lab cognitive function assessment and eliminate a great many of the drawbacks of a functional MRI. fNIR optical imaging technology measures neural activity and hemodynamic response in the prefrontal cortex. Subjects wear an fNIR sensor (IR light sources […]

View All
Latest News

NEWS CITATIONS – Heart Rate & Heart Rate Variability, Wingate Anaerobic Test, and Neuroscience of Music

BIOPAC provides software and hardware that allow for research teams to record and analyze respiration activity in physiological experimentation. Here are a few notable studies covering HRV, EMG, and EEG measurement. HRV (Jing Wei, Hong Luo, Si J. Wu, Paul P. Zheng, Genyue Fu, and Kang Lee) Humans encounter various stressful situations everyday at work, home, and school. Such stress when experienced at high degrees and/or […]

What makes you laugh the most?

Alan Macy Answers Proust Questionnaire BIOPAC co-founder and R&D Director, Alan Macy, was interviewed recently by Roger Durling for the Santa Barbara Independent. The article, “The S.B. Questionnaire: Alan Macy,” provided a background excerpt of the Santa Barbara entrepreneurs’ past as well as his answers to the Proust Questionnaire. The biomedical technology developed at BIOPAC reflects an ambitious […]

Read All
Request a Demonstration
Request a Demonstration