Ground vs. reference for EEG recording

The terminology for electrical connections to human subjects is not universally agreed upon. “Ground” and “reference” are often used interchangeably. This is not a good practice as there are different purposes for different connections.

Ground and reference are better understood as two different types from the perspective of both physiology and amplifier design. BIOPAC uses a relatively pervasive terminology here.

Alpha Rhythms in the Occipital Lobe Clip Lead White 1 m Shielded Back

In a unipolar setup, there will be one reference electrode, and the potential difference between this electrode (connected to Vin-) and every other electrode on the head (connected to Vin+ on separate amplifiers) will be measured.

For bipolar recordings, each active (Vin+) electrode will have its own reference (Vin-) connection.

BIOPAC Systems use a common ground connection for all amplifiers that are electrically connected to each other (i.e., part of the same MP system and not passing through a wireless stage as the BioNomadix modules do). In these cases, there should not be more than one ground connection to the subject unless all but one of the ground connections are through AC coupled leads. For instance, recordings made through a single MP36 system should have only one ground connection to the subject. Similarly, recordings made through EEG100C or EEG100C-MRI amplifiers all connected to a single MP150 system should have only one ground connection (total!) to the subject. BioNomadix wireless physiology data systems necessarily have a separate ground for each transmitter (note that each transmitter has two amplifiers on it). In this case, each transmitter should have one ground connection to the subject even if all transmitters communicate with the same MP150 system.

 

modified 12/31/14

Associated Applications

Spotlight On

Noninvasive Blood Pressure Amp with Hemodynamics

Continuous…Easy to use…Reproducible & Validated The NIBP100D-HD noninvasive blood pressure system provides a continuous, beat-to-beat, blood pressure signal recorded from the fingers of a subject. Simple and noninvasive—finger sensor provides accurate & immediate feedback: arterial BP, cardiac output, fluid and hemodynamic status. Measure noninvasive BP parameters (BP, sBP, dBP, mBP, and PR) plus hemodynamic parameters (PPV, […]

View All
Latest News

New Citations | Psychophysiology, Chewing Study, Pulse Wave Velocity

Recent studies in Psychophysiology, EMG, ECG and PPG measurement utilizing BIOPAC software and hardware: Fight or Flight (Reflex Response Time) “Fight or flight” is a physiological response that occurs due to a perceived threat to survival. It can physiologically be observed through an increase in heart rate, respiratory rate, blood pressure, oxygen availability, and other mechanisms. In […]

New Citations | Heart Rate & Heart Rate Variability, Wingate Anaerobic Test, and Neuroscience of Music

HRV, EMG, and EEG studies using BIOPAC hardware and software include: DON’T STRESS Humans encounter various stressful situations everyday at work, home, and school. Such stress when experienced at high degrees and/or for a long duration of time could lead to cardiovascular diseases, cognitive dysfunctions, and psychological disorders. This study examined the validity of a novel physiological measurement […]

Read All
Request a Demonstration
Request a Demonstration