Auditory Brainstem Response (ABR)

MP Devices can be used for a wide variety of Brainstem Evoked Potential (BEP), Brainstem Evoked Response (BER), Brainstem Auditory Evoked Potential (BAEP), Brainstem Auditory Evoked Response (BAER), Brainstem Evoked Response Audiometry, and Brainstem Evoked Response (BSER) tests. The MP device is capable of delivering auditory stimulation and recording and averaging the response data. The system can also interface with a wide variety of stimulus presentation systems to deliver auditory waveforms and the averaging mode can handle oddball paradigms for Middle Latency Auditory Evoked Potential (MAEP) and Auditory Middle Latency Response (MLR) tests. The averaging window can be adjusted to include the time period of interest for the particular response being studied.

For ABR studies the STM100C Stimulator module should interface with the OUT101 Ear Tubephone to deliver clicks and tones to the subject. Use the ERS100C to record the evoked response from the subject and set AcqKnowledge to average the signal.

 

See More...

Details

This is one of many ADVANCED FEATURES for the selected Application. Scroll down for hardware options.

Record neuro-physiological responses along the auditory pathway resulting from auditory stimulation. When performing an ABR, using the MP150, it’s usually best to record at least two channels of data. One channel is the output from the ERS100C amplifier connected to the subject and the other channel is the stimulus level which drives the OUT101 Tubephone or other auditory stimulator.

Latency values, which are not possible to achieve, are reflected as an adjustment message (generated by the AcqKnowledge software) when the acquisition is first performed. The message informs the user as to the options to continue with the acquisition, given the acquisition parameters. If the specified Latency is possible, then the adjustment message will not appear and the acquisition will start automatically when the START button is pressed.

The fastest repetition rate (smallest latency between clicks) will be a function of the following variables:

1) Acquisition Length: the longer the length, the higher the latency
2) Sampling Rate: the higher the sampling rate, the higher the latency
3) Number of Channels: the higher the number of recorded and averaged channels,
the higher the latency
4) Speed of Computer: the slower the computer speed, the higher the latency

Calibration for Auditory Brainstem Response Studies

To calibrate the OUT101 Tubephone, use an Etymotic ER-7C Probe Microphone—this microphone provides a calibrated output voltage which is a function of applied Sound Pressure Level (SPL). The sensitivity is 50 mV/Pascal (-46 dB re: 1 V/uBar): 0 dB SPL = 0 dBuV.

Place the Probe Microphone insert tube in the auditory canal prior to the insertion of the OUT101 foam tip. The OUT101 Tubephone sound delivery tube and the Probe Microphone sound input tube will then be exposed to the same auditory chamber. Accordingly, the SPL is recorded, via the Probe Microphone, simultaneously with applied auditory stimulus from the OUT101 Tubephone.

See also

 105c – ABR Calibration for MP150 with AcqKnowledge 4.1       249 – Online Averaging Latency Parameters

Support

Application Notes

Knowledge Base

Hardware Packages   |   Auditory Brainstem Response (ABR)

Hardware Bundles are complete solutions for the specified application. Choose your preferred platform and bundle, then click "Request Pricing" to request an estimate, add/remove items, or complete purchase. If you have questions about specific items, click through to the product web page for details and specifications, or contact your Local Sales contact.

Wired

MRI

Animal

Spotlight On
Ventilator Validation Kit

Ventilator Validation and Test Equipment

The Ventilator Validation Kit supports new prototype development, existing design validation for production testing, mobile field testing and verification, and other uses where accurate pressure and flow measurements are critical. The Ventilator Validation Kit provides ventilator manufacturers and prototype developers with a complete solution for validating new medical/hospital-grade ventilator products intended for human use. The VVK100-SYS includes […]

View All
Latest News

BIOPAC Systems Helps STARK Industries Launch VITAL Ventilator for NASA JPL

BIOPAC’s ventilator testing equipment and expertise supports STARK Industries’ effort to manufacture, validate, and test NASA JPL’s VITAL Ventilators Goleta, California – June 16, 2020 BIOPAC Systems, Inc. announced a partnership with STARK Industries, LLC to provide ventilator validation and testing equipment for the mass production of ventilators. STARK Industries was awarded a global license […]

New Citations | Go Inside with Imaging Solutions from BIOPAC!

Physiological Lung Testing in the MRI to Find Damage from Military Deployments Our military veterans who served overseas in the middle east often come home with scars both seen and unseen. For some vets, unfortunately, these scars can take the form of lung damage sustained due to sandstorms, burn pits, and other high toxicity situations apparent […]

Read All
Request a Demonstration
Request a Demonstration