PRO software will control the pulse width, frequency and output (single pulse or series) options.
The BSL PRO software will automatically note any change to the stimulator parameters (e.g. pulse amplitude, width, frequency) and will insert an event marker at the point of change and include text to describe the change. You can set the Stimulator to start automatically when you start the recording, or manually at any point during the recording. You can display and monitor stimulus data independently of the sample rate or stimulus frequency.
The stimulator can be used for the following physiological measurement protocols:
Twitch sub-threshold
Twitch threshold
Maximum twitch responses
Single twitch, summation
Tetanic contraction
Muscle tension/length versus force
Fatigue
Nerve conduction velocity
Stimulation electrodes:
HSTM01Human-safe Handheld Stimulation Electrode
ELSTM2Needle Electrode
WARNING Like all stimulating devices, the BSLSTM stimulator can be dangerous--please follow all stimulation safety guidelines!
When powering ON the BSLSTM stimulator and/or the MP36/35 unit, follow this procedure for maximum safety: Connect the Trigger and Reference Output cables from the BSLSTM to the MP36/35 unit.
Turn the MP36/35 unit ON.
Launch the BSL PRO Software.
Click on the MP36/35 menu and select Output Control>Stimulator.
Turn the Level knob on the BSLSTM to 0 volts (fully counter-clockwise).
Flip the BSLSTM Power switch ON.
Attach the Output from yhe BSLSTM to the subject.
Turn the Level knob to the desired initial output level.
Click on the "START" button in the BSL PRO software to start acquiring data -- Preferences may be set to start stimulation with start of acquisition or with the switch in the Stimulator output Control window.
Incrementally adjust the Level for the desired Output.
When you have completed recording and taken the required measurements, turn the Level back to 0 volts.