HEAD FOR TSD137A AND SS45L 12ML/SEC - RX137A

For Estimate or Purchase
Request Information
Head for TSD137A and SS45L 12ml/sec

The RX137A is a replacement airflow head for the TSD137A or SS45L pneumotach transducer. RX137 series heads can be mixed and matched with any of the TSD137 series or SS45L-SS52L series pneumotachs. A wide range in flows can be accommodated by switching one head for another when using a single pneumotach. RX137 heads connect to the TSD160A or SS40L differential pressure transducer via standard 4mm ID tubing. Multiple RX137 heads help eliminate equipment downtime during cleaning procedures.

Cleaning & Disinfection:

IMPORTANT:

  • RX137 Series airflow heads are manufactured with a very thin layer of synthetic resin, so they should never be cleaned with an organic solvent. We recommend cleaners such as Hydro-Merfen at the concentration used for medical material, or Gluterex.
  • Before using the airflow head, be sure it is dry.
  • Never heat the airflow head higher than 50° C.
  1. Submerge the airflow head in a disinfectant solution for approximately one hour.
  2. Rinse the airflow head with tap water.
  3. Rinse the airflow head with distilled or de-mineralized water.
  4. Use compressed air or another compressed gas [pressure up to 5 kg / cm2 (5 bar)] to drive any remaining water out of the airflow head.
  5. Allow the airflow head to dry completely in ambient air (or continue using compressed air if time requires it).

 