The DA100C is a very low noise and drift differential bridge amplifier that will operate with a wide variety of transducers, including: Pressure, force, strain, displacement, thermocouples, piezo sensors, light sensors, microphones and nearly any other active or passive sensor or transducer. The DA100C provides variable gain settings, offset control for baseline zeroing and adjustable voltage references for activating passive transducers.
See the TCI Seriesof more than 15 transducer connector interface options, including a custom interface kit.
Gain: 50, 200, 1000, 5000 Output Range: ±10 V (analog) Frequency Response Maximum bandwidth (DC-5,000 Hz) Low Pass Filter: 10 Hz, 300 Hz, 5000 Hz High Pass Filter: DC, 0.05 Hz (internal, optionally selected single pole filter) Input Voltage (max): ±200 mV (protected) Noise Voltage: 0.11µV rms – (0.05-10 Hz) Temperature Drift: 0.3µV/°C Z (Differential input): 2 M ohm CMRR: 90 dB min CMIV—referenced to Amplifier ground: ±10 V Mains ground: ±1500 VDC Voltage Reference: -10 to +10 V infinitely adjustable @ 20ma (max) (preset to 2 volts excitation) Signal Source: Variety of transducers Input Voltage Range Gain Vin (mV) 50 ±200 200 ±50 1000 ±10 5000 ±2 Weight: 350 grams Dimensions (WxDxH): 4 cm x 11 cm x 19 cm