The SS3LA transducer connects to a single MP3X input channel to record electrodermal activity (skin conductance or, with proper setup
*, skin resistance). Two Ag-AgCl electrodes are mounted in individual, ergonomically designed, polyurethane housings for improved contact; they attach to the fingers by a Velcro strap or can be taped to any other part of the body. The electrodes have a 6 mm diameter contact area with a 1.6 mm cavity to accommodate electrode gel. The non-polarizable electrodes are shielded to minimize noise interference and improve recordings.
BSL
PRO software includes two EDA presets:
Electrodermal Activity (EDA), 0 - 35 Hz; requires calibration—see
BSL Hardware Guide (under the Help menu or online) for details
Electrodermal Activity (EDA) Change; no calibration required
See the
BSL Hardware Guide (under the Help menu or online) for details on how to verify the Gain setting of the SS3LA.
*Skin resistance setup: Use an Expression calculation channel to take the reciprocal of conductance and then apply proper scaling. See the BSL
PRO Software Guide (under the Help menu or online) for setup details.
Usage Recommendations
There must be good electrical connections between the skin and the electrodes for EDA to work properly. When using GEL101 isotonic gel it is important that the gel has a chance to be absorbed and make good contact before recording begins. Accordingly:
Apply GEL101 to the skin at the point of electrode contact and rub it in.
Fill the SS3LA electrode cavity with GEL101.
Attach the SS3LA electrode to the subject.
Wait 5 minutes (minimum) before starting to record data.
See the SS57L EDA Lead for a disposable electrode option