Use the OUT101 tubephone to deliver clicks and tones in auditory evoked response applications (i.e. ABR). The Tubephone design consists of a monaural acoustic transducer attached to a short, flexible, plastic tube, which fits into the subjects ear with the aid of a foam tip. Use of the Tubephone reduces ambient noise and bone conduction problems, which can interfere with auditory response recordings. Furthermore, because the Tubephone provides a 1 msec acoustic signal delay (due to plastic tube), it automatically separates true response from electromagnetic artifact resulting from speaker activation.
MP150/100 System interface:
STM100C: connects directly to stimulator model
MP36 and MP36R System interface options:
BSL System stimulator (model BSLSTM): use BSLCBL6 and Radio Shack P/N 274-047 ¼" to 1/8" phono adapter
MP36/36R Analog Out port—no stimulator required!—use OUT3 plus BSLCBL6 and Radio Shack P/N 274-047 ¼" to 1/8" phono adapter
MP36/36R Headphone port—no stimulator required!—use Radio Shack P/N 274-047 ¼" to 1/8" phono adapter; note—volume may not reach the same levels as the Analog Out port
Calibration for Auditory Brainstem Response Studies
To calibrate the OUT101 Tubephone, use an Etymotic ER-7C Probe Microphone—this microphone provides a calibrated output voltage which is a function of applied Sound Pressure Level (SPL). The sensitivity is 50 mV/Pascal (-46 dB re: 1 V/uBar): 0 dB SPL = 0 dBuV.
Place the Probe Microphone insert tube in the auditory canal prior to the insertion of the OUT101 foam tip. The OUT101 Tubephone sound delivery tube and the Probe Microphone sound input tube will then be exposed to the same auditory chamber. Accordingly, the SPL is recorded, via the Probe Microphone, simultaneously with applied auditory stimulus from the OUT101 Tubephone.