High-accuracy fiber optic temperature probes
Rectal probe: 420 µm OD of Polyimide tubing; 900 µm OD tight buffer PVC, 8 m, MRI compatible
Use FO probes with the stand alone FOTS100 Fiber Optic Temperature System.
To interface with a BIOPAC MP System, add a cable interface (separate purchase):
Applications Magnetic Resonance Imaging - MRI Compatible
|
| |
|
Temperature range: 0 °C to +85 °C (other ranges AUR)
Response Time: 250 ms and better
Temperature operating & calibrated range: 20°C to 45 °C (other ranges AUR)
Accuracy: ±0.2°C (Total accuracy over the calibrated range including both signal conditioner and sensor errors)
Resolution: 0.05 °C
Operating humidity range : 0-100 %
MRI/EMI/RFI susceptibility: Complete immunity
Calibration: NIST traceable
Optical connector: ST standard
Cable sheathing: 420 µm OD of Polyimide tubing; 900 µm OD tight buffer PVC
Cable length: 8 m
Signal conditioner compatibility: FOTS100 system
OTG-MP-62ST-8m |
|
|
|
| |
|
|