The TSD125 series force transducers interface with the DA100C general-purpose transducer amplifier. The TSD125 series force transducers have been designed for in-vitro tissue bath studies, in-vivo force measurements and other applications where low noise, accuracy and repeatability are critical. The most sensitive units are ideal for use with small preparations such as aortic rings. The TSD125 series utilizes a precision strain gauge element, which has superb low noise characteristics and excellent linearity. The transducers come with a versatile mounting rod assembly to support a variety of orientations.
When fine tension and position adjustments are required, clamp the mounting rod to an HDW100A Force Transducer Tension Adjuster (not included).
FULL SCALE
10Hz Low Pass Filter
1 Hz Low Pass Filter
DEVICE
RANGE (FSR)
Noise (rms)
Noise (rms)
TSD125C
50 gram
2.5 mg
1 mg
TSD125D
100 gram
5 mg
2 mg
TSD125E
200 gram
10 mg
4 mg
TSD125F
500 gram
25 mg
10 mg
See note*
Full Scale Output:
1mV/V (normalized to 1V excitation)
Nonlinearity:
<0.025% FSR
Hysteresis:
<0.05% FSR
Nonrepeatability:
<0.05% FSR
30 Minute Creep:
<0.05% FSR
Temperature Range:
-10C to 70°C
Thermal Zero Shift:
<0.03% FSR/°C
Thermal Range Shift:
<0.03% Reading/°C
Recommended Excitation:
10 VDC (max)
Weight:
250 grams
Mounting Rod:
9.5mm (dia)- variable orientation
Cable Length:
3 meters
Dimension:100mm (long) x 19mm (wide)x 25mm (thick)
*The table shows typical noise levels with 10Hz and 1Hz Low Pass Filters with 10 volts excitation.